Seite 1 - Oxygen Saturation Monitor

Oxygen Saturation MonitorPULSOX™-2Instruction Manual9222-1731-14_P-2.indd i9222-1731-14_P-2.indd i 09.8.20 11:03:21 AM09.8.20 11:03:21 AM

Seite 2 - Authorized Standards

8MeasurementsMeasurements must be taken with the index fi nger placed in the fi nger holder. This instrument must be held tightly. Put the thumb only on

Seite 3 - SAFETY PRECAUTIONS

9Safety Notes This finger holder is designed only for use on the fingers of adults, so it should not be used on any other body parts. Never secure the

Seite 4 - 1. BEFORE OPERATION

10Ending Measurements1. Open the fi nger holder and remove the fi nger.• Measurement will stop automatically, and the display will go off after 10 seco

Seite 5 - 2. NAMES OF PARTS

11How to Read the Measurement ValuesPulse level meter: Indicates the pulse level in 8 levels (0 to 10%, full scale if 10% or higher). The pulse level

Seite 6 - 3. NOTES ON USE

12Back LightThe backlight will light up automatically when the surroundings be-come dark. When backlight isn’t turned by the bright surroundings, cove

Seite 7

13Error Messages Messages relating to connection of the fi nger holder and attachment to the patientError MessagesThe Measure-ment Value andflash alt

Seite 8 - (1) Installing Batteries

14Error MessagesThe Measure-ment Value andCause Solution• Motion artifact• The pulse rate ex-ceeds the upper limit (250 bpm) of the measurement range

Seite 9

155. TROUBLESHOOTING Main Unit• Display disap-peared in the middle of measurements.• No display ap-pears when the finger holder was opened and the me

Seite 10 - Measurements

166. In Case of Malfunction• If a malfunction occurs, contact the nearest authorized service facility. Never disassemble the unit or attempt to rep

Seite 11 - Notes on Measurements

1710. SPECIFICATIONSPULSOX-2 Dual-wavelength pulse-type oximeter Functions:• Measuring range: SpO2: 0 to 100% Pulse rate: 20 to 250 bpm• Accura

Seite 12 - Ending Measurements

This indicates text consisting of a warning or precaution relating to safety. Please read the text carefully and use the equipment safely.For North Am

Seite 13

1811. REFERENCEMeasurement PrincipleOxygen Saturation Monitor PULSOX-2 is a photometric instrument that non-invasively and continuously measures the o

Seite 14 - Backlight sensor window

19Relation between Oxygen Saturation and Partial PressureThe relation between oxygen saturation (SpO2, %) and oxygen partial pressure (PaO2, Torr or

Seite 15 - Error Messages

20Guidance and manufactures declaration - electromagnetic emissionsGuidance and manufactures declaration - electromagnetic immunityThe Model PULSOX-

Seite 16

21Guidance and manufactures declaration - electromagnetic immunityThe Model PULSOX-2 is intended for use in the electromagnetic environment specified

Seite 17 - 5. TROUBLESHOOTING

9222-1731-163-91, Daisen-nishimachi, Sakai-ku, Sakai-shi, Osaka 590-8551, JapanBDEAPX Printed in Japan© 2002-2013 KONICA MINOLTA, INC.

Seite 18 - 9. Cleaning

1SAFETY PRECAUTIONSTo ensure correct use of this instrument, read the following points carefully and adhere to them. After you have read th

Seite 19 - 10. SPECIFICATIONS

21. BEFORE OPERATIONSafety Notes CAUTION: U.S. Federal law restricts this device to sale by or on the order of a physician. The PULSOX-2 is a medic

Seite 20 - Red light Infrared light

32. NAMES OF PARTS Main unitBattery cover: Open or close this cover when replacing the batteries with new ones.Battery cover lock: It is used to loc

Seite 21 - Oxygen Partial Pressure SpO

43. NOTES ON USE• This instrument should be used in areas with an ambient temperature of 0 to 40°C (32 to 104°F) and a relative humidity of 30 to 95

Seite 22 - EMC Declaration

5This equipment has been tested and found to comply with the limits for medical devices to the IEC 60601-1-2:2001, Medical Device Directive 93/42/EEC

Seite 23 - E㲋1

64. OPERATING METHODPreparations(1) Installing Batteries(Two AAA-size batteries are required.)1. Push the battery cover, and unlock the battery cover

Seite 24 - 9222-1731-16

75. Close the battery cover lock until it clicks. • The batteries will last approxi-mately 80 hours.• The battery mark will begin to blink when th